User Information
Introduction to Headspace and to the G1888 Network Headspace Sampler
Filling the sample loop
The vent valve opens and the headspace gas fills the sample
loop. It then vents to atmospheric pressure through the vent
outlet. A short loop–fill time (2 to 5 seconds) may leave
pressure in the loop equilibrated to the vial pressure and allow
more analyte to be injected. A longer loop–fill time
(10 to 15 seconds) equilibrates the loop pressure to atmospheric
pressure. See
Figure 9
Loop fill. S1 is closed. S2 is open to fill the loop.
Carrier gas flow
Mass
controller
Pressure
transducer
Vent
Sample
loop
To GC
Auxiliary gas flow
for vial pressurization
Pressure
regulator
Pressure
transducer
Flow restrictor
S1
S2
Needle
Headspace vial
Open
Closed
Carrier gas
Auxiliary gas
1 2
3
4
5
6
flow
Toggle
switch
Toggle
switch
medium_standard.book Page 38 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...