Troubleshooting and Error Messages
User Information
171
Carryover in air or solvent blanks
In theory a run containing a blank made after a run with a
sample has no peaks in the chromatogram except for the
analytes from the lab air or solvent that is trapped in the sample
vial. Carryover results when sample condenses on the flow path
or is trapped in unswept area from a sample injection. Run a
blank to see small quantities of these analytes.
If the amount of carryover in a series of solvent blanks
decreases by a factor of 5 to 10 for each run, the carryover is
probably due to an unswept area. Check the following:
•
There is flow from the GC inlet to sweep the inlet weldment.
•
Check the transfer line connection to the GC inlet.
If the amount of carryover in a series of solvent blanks does not
decay for each run, the carryover is probably due to analytes
adsorbed or condensed on the flow path or in the inlet. Check
the following:
•
Poor choice of heated zone temperature settings. See
•
Poor choice of vial pressurization pressure or time.
See
•
Flow path or GC inlet contamination.
•
Poor choice of inlet.
medium_standard.book Page 171 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...