The Keypad
User Information
57
Vial Parameters
Description
Select
Vial Parameters
from the Active Method menu to
specify the vial(s) analyzed during a method and access the
shake function.
First Vial
Designates the first vial (1–70) analyzed by the
current method.
Last Vial
Designates the last vial (1–70) analyzed by the
current method. The last vial number cannot be lower than the
first vial number. The Headspace Sampler will give an error
message and you will need to re-enter the vial number.
Shake
Shake specifies the agitation level of the oven carousel
during vial equilibration. Shaking the vial may shorten the
needed vial equilibration time.
There are three shake settings: off, low, and high. Specify the
level desired using the cursor keys followed by
Enter
. Low is
sufficient for most applications.
First Vial
1
Last Vial
22
SET
VIAL PARAM.
Shake
Off
Not Ready
medium_standard.book Page 57 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...