132
User Information
Maintaining the Instrument
Changing the Sampling Probe
Follow the steps in
to change the sampling probe.
Table 11
Changing the sampling probe
Step
Action
Notes
1
Remove the loop cover. Follow the
steps on
.
2
Remove the heater block covering
the sampling probe.
a
Loosen the Phillips screw shown
below.
b
Lift out the heater block.
Screw
Heater block
medium_standard.book Page 132 Tuesday, February 17, 2004 10:14 AM