Installing the flow manifold
7
Installing the flow manifold
This section describes installation of the current version of the pneumatics
module. For information on the earlier version, see the Appendix.
Caution
It is not necessary and not advisable to separate detectors from their
pneumatics modules. Doing so can cause leaks. Although handling the
detector and pneumatics module as a unit is awkward, it can be managed.
1.
Locate the correct slot for the pneumatics module.
•
Detector in the front position.
Use the second slot from the left, as
viewed from the back of the GC.
•
Detector in the back position.
Use the third slot from the left, as
viewed from the back of the GC.
Caution
Hold the manifold by its support bracket to avoid damaging the components.
2.
Route the ribbon cable behind the manifold assembly as shown in
Figure 1. Slide the manifold and bracket assembly into the slot until the
bracket seats flush against the end of the rails. See Figure 2.
Figure 1.
Routing the ribbon cable
Back view of manifold
Manifold installed with cable routed to left