Theory of Operation
Firmware Updates
Firmware updates can be done using your user interface:
•
handheld control module with files from a PC-card or
•
Agilent ChemStation with files from floppy disk
The file naming conventions are:
xxxx-vvv.DLB, where
xxxx is the product number, e.g. 1367A for the well-plate sampler), and
vvv is the revision number, for example 380 is revision 3.80
For instructions refer to your user interface.
N O T E
Update of main system can be done in the resident system only.
Update of the resident system can be done in the main system only.
Figure 40
Firmware Update Mechanism
Resident System
Main System
Resident FW update
Main FW update