140
1100 Series WS MWS Reference Manual
4
Repairing the Sampler
Installing the injection valve assembly
1
Feed the cable of the new valve through to the SUD board.
2
Connect the valve cable to the connector labelled “valve assy”.
3
Install and tighten the three screws which fix the injection valve assembly.
4
Install the sampling unit, (see
“Installing the sampling unit"
5
Install the sample transport (see
“Installing the sample transport"
6
Install the top cover, top plate, and foam (see
7
Replace the plate tray in the tray base.
8
Turn ON the sampler, close the front door.
9
The system will reset automatically.
10
In the user interface perform a “Sample Transport Self Alignment”.
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......