1200 Series Handheld Control Module Users’s Guide
65
Using the Pump
3
Purging Procedure
Purging Procedure
1
Ensure that an outlet tube is connected from the purge valve to a waste
solvent bottle.
2
Open the purge valve.
3
Enter the
Analysis
screen.
4
Set the purge flow rate in the regular flow rate section of the
Analysis
screen, for example, to 5 ml/min.
5
Set the first channel to be purged to 100%. We recommend you start with
the organic solvent channel first. Channel A will automatically be set to
100% when all other channels are set to zero or OFF.
6
Turn on the pump using the
On/Off
button [
F7
] in the
Analysis
screen.
7
Wait until a continuous stream of solvent comes out of the outlet tube from
the purge valve.
8
Turn off the pump using the
On/Off
button [
F7
] in the
Analysis
screen and
then close the purge valve.
9
Repeat the procedure for the other channels you need to purge.
N O T E
If a flow rate of more than 5 ml/min is used on a quaternary pump, redefine the upper
pressure limit to 200 bar (Accessible from the Analysis screen, press the Settings button
[F1], choose pump from the list and press the Enter key. Then choose More ...)
N O T E
The channels of a pumping system are named A, B, C and D (depending on the pump type).
%A is automatically calculated by 100% - (%B + %C + %D). If no values for %B, %C and %D
are entered, %A is always 100%. To purge the pump you have to go through steps
to
individually for each of the channels, setting the composition to 100% for the
channel to be purged.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......