152
1200 Series Handheld Control Module Users’s Guide
12
Running an Injector Program
Entering Injector Program Settings
Injector Program Lines
describes the program lines which will complete the required
derivatization reaction.
Table 6
Injector Program Lines
Line
Function
Parameters
Comment
1
DRAW
2 µl (AMOUNT) from Air
(SOURCE)
Draw 2 µl from air into the needle
capillary, to separate the mobile phase
sitting in the capillary from the sample.
2
DRAW
1 µl (AMOUNT) from vial
1(SOURCE)
Draw 1µl from derivatization reagent
from vial #1 in the autosampler tray into
the needle capillary.
3
WASH
in vial 2 (VIAL), 1 time
(Cycles)
Wash the needle tip in the wash vial #2.
4
DRAW
2 µl (AMOUNT) from sample
(SOURCE)
Add 2 µl of the sample in the
autosampler tray to the derivatization
volume already in the needle capillary.
5
WASH
in vial 2 (VIAL), 1 time
(CYCLE)
Wash the needle tip in the wash vial #2.
6
DRAW
1 µl (AMOUNT) from vial 1
(SOURCE)
Add 1 µl of derivatization reagent from
vial #1 in the autosampler tray to the
derivatization reagent and sample
already in the needle capillary.
7
MIX
6 µl (AMOUNT) in seat
(SOURCE), at 500 µl/min
(SPEED), 8 times (REPEAT)
Mix 6 µl with a repetition of 8 times at a
speed of 500 µl/minute in the seat.
8
WAIT
0.5 minutes (WAIT)
Wait for 0.5 minute for the mixture to
completely react.
9
INJECT
Inject the mixture and start the analysis.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......