116
1200 Series Handheld Control Module Users’s Guide
9
Running an Isocratic Analysis
Entering Settings
Entering Settings
To set up the isocratic analysis you will set the LC system settings to default
and then modify selected settings, the other settings will remain with their
default values. You will then save these settings to a method called ISO.
1
Enter the
Analysis
screen.
2
Set the vial range as 1 to 1.
3
Set the number of injections to 1.
4
Press Enter to skip the method name section.
5
Select the
Settings
button [
F1
].
6
Select LC System from the menu.
7
Select the
Default
button [
F7
] and select the
Yes
button to load defaults.
8
Press
Esc
.
9
Select the
Settings
button [
F1
].
10
Select the pump from the menu (either ISO, BIN or QUAT).
11
Press the
Pressure
button [
F3
].
12
Enter 400 as the upper pressure limit.
13
Select the
Done
button [
F6
].
14
Enter the following values: %B 65, (%C OFF, %D OFF for quarternary pump.
If you have a binary pump, set %B to 65.).
15
Set the Flow to 1,5 ml/min.
16
Confirm by selecting “
Done
” to enter the
Analysis
screen
17
Enter the values shown in
in the
Analysis
screen.
Table 4
Values in Analysis Screen
Setting
Value
Detection wavelength
254 nm
Injection volume
1 µl
Stoptime
6 min
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......