Wavelength holmium check failed
Error ID: 7318
The holmium oxide test in the detector has failed. During the holmium
test, the detector moves the holmium filter into the light path, and
compares the measured absorbance maxima of the holmium oxide filter
with expected maxima. If the measured maxima are outside the limits, the
error message is generated.
Grating or Filter Motor Errors
Error ID: Grating: 7800, 7801, 7802, 7803, 7804, 7805, 7806, 7808, 7809;
Filter: 7810, 7811, 7812, 7813, 7814, 7815, 7816
The motor test has failed.
During the motor tests, the detector moves the motor to the end position
while monitoring the end- position sensor. If the end position is not found,
the error message is generated.
Probable cause
Suggested actions
1
Misaligned/defective grating assembly.
•
Ensure the flow cell is inserted correctly,
and is free from contamination (cell
windows, buffers, and so on).
•
Run the filter-motor test to determine if the
filter motor assembly is defective. If
defective, please contact your Agilent
service representative.
•
Run the grating-motor test to determine if
the grating assembly is defective. If
defective, please contact your Agilent
service representative.
Test 0 Failed:
Filter motor.
Test 1 Failed:
Grating motor.
Содержание G1314B
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Страница 30: ...30 Agilent 1260 Infinity VWD User Manual 2 Site Requirements and Specifications Performance Specifications...
Страница 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Страница 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Страница 190: ...190 Agilent 1260 Infinity VWD User Manual 12 Hardware Information Early Maintenance Feedback EMF...
Страница 207: ......