81
Troubleshooting and Test Functions
Change Needle
Change Needle
WA R NI N G
For needle exchange, the needle arm moves down automatically when
the front cover is removed. To avoid personal injury, keep fingers away
from the needle area during needle movement.
The change-needle/seat function moves the safety flap out of position, and
positions the needle for easy exchange and alignment of the needle and
needle seat.
User Interface
The commands for the ChemStation and Control Module (different names for
the commands in the Control Module are shown in brackets) are:
N O T E
The autosampler front cover must be in place when “Start” and “End” are
selected.
Start (Change)
Moves the safety flap away from the needle, and positions
the needle approximately 15 mm above the needle seat.
Needle Up (Up
Arrow)
Press function key couple of times to move the needle arm
up in 2 mm steps.
Needle Down
(Down Arrow)
Press function key couple of times to move the needle arm
down in 2 mm steps. The lowest position (“end position”)
is used to align the needle at the correct position in the
needle seat.
End (Done)
Completes the procedure by moving the gripper arm to the
home position, and releasing the safety flap.
Using the Change Needle Function
1
Ensure the front cover is installed.
2
Select “Start” (“Change”) to move the needle arm to the maintenance position.
Содержание G1313A
Страница 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Страница 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 12: ...12 Contents ...
Страница 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Страница 49: ...3 3 Troubleshooting and Test Functions The modules built in troubleshooting and test functions ...
Страница 93: ...93 Troubleshooting and Test Functions Step Commands ...
Страница 159: ...5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials ...
Страница 160: ...160 Parts and Materials Main Assemblies Main Assemblies Figure 19 Autosampler Main Assemblies 5 3 4 2 1 9 8 7 6 ...
Страница 166: ...166 Parts and Materials Analytical Head Assembly Figure 21 Analytical Head Assembly 2 3 4 5 6 1 ...
Страница 191: ...191 Parts and Materials Remote Cables NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector ...
Страница 216: ...216 Introduction to the Autosampler Electrical Connections ...
Страница 217: ...7 7 Theory of Operation Theory of operation of mechanical hardware electronics and instrument interfaces ...
Страница 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Страница 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Страница 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Страница 269: ...9 9 Specifications Performance specifications of the Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 274: ...274 Specifications Performance Specifications ...
Страница 275: ...10 Legal Safety and Warranty Information ...
Страница 286: ...286 Agilent Technologies on Internet ...
Страница 290: ...290 Index ...
Страница 291: ......