![Agilent Technologies FRG-700 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/agilent-technologies/frg-700/frg-700_short-operating-instructions_2868123010.webp)
Inverted Magnetron Pirani Gauge FRG-700 FRG-702
10
2
Technical Data
Measuring range (air, N
2
)
5×10
-9
… 1000 mbar
Accuracy
±30% (in the range
1×10
-8
… 100 mbar)
Repeatability
±5% (in the range
1×10
-8
… 100 mbar)
Output signal (measuring signal)
Voltage range
0
… +10.5 V
Measurement range
1.82
… 8.6 V
Voltage vs. pressure
logarithmic , 0.6 V/decade
Error signal
<0.5 V no supply
>9.5 V Pirani sensor de-
fective (filament rupture)
Output impedance
2×10
Minimum loaded impedance
10 k
, short-circuit proof
Response time
p > 10
-6
mbar
(pressure dependent)
<10 ms
p = 10
-8
mbar
1000 ms
Identification gauge
85 k
, referenced to
supply common
Status
pin 6
p > 10
-2
mbar
Pirani-only mode
Low = 0 V
p < 10
-2
mbar
Cold cathode not ignited
Pirani-only mode
Low = 0 V
p < 10
-2
mbar
Cold cathode ignited
Combined Pirani / cold
cathode mode
High = 15
… 30 VDC
LED
High voltage on (LED on)
Содержание FRG-700
Страница 1: ......
Страница 2: ......
Страница 3: ...This page is intentionally left blank ...
Страница 32: ...Inverted Magnetron Pirani Gauge FRG 700 FRG 702 32 Notes ...
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 36: ...t q ma 7 4 e 1 ...