Flow Tracker Series 1000/2000
Flow Tracker 1000/2000 Flow Measurement Specifications
provides flow measurement specifications for the Flow
Tracker 1000 and Flow Tracker 2000 flowmeters.
Table 6
Flow measurement specifications
Flow range for measurement
0–500 mL/min
Accuracy
(over 5–500 mL/min)
Flow:
±
2% of reading,
±
0.2 mL/min
(
±
5% from 0–5mL/min)
*
* Median reading using test procedure. Test procedure can be found at: www.agilent.com/chem/supplies under the gas
management section
Temp:
±
0.5 °C
Press:
±
0.5% of reading
Sensor type
Flow:
Solid-state piezoresistive differential pressure sensor
with measurement in the laminar flow region. No
heated elements
Temp:
Integrated circuit absolute temperature sensor. No
RTD’s
Press:
Solid state absolute pressure sensor
Calibration
NIST traceable multi-point calibration certificate included
Calibrated gases
Nitrogen, hydrogen, helium, air, carbon dioxide, methane,
95% argon / 5% methane blend
Modes
Flow, gas chromatography, leak detection (Flow Tracker 2000 only)
Display
Dynamically labeled multi-function 7 line LCD with 99 characters
Power
Six AA batteries (included) or optional universal AC/DC Adapter
(Part number 5183-4781)
Dimensions
8.3-inch (h) x 3.75-inch (w) x 1.9-inch (d)
Certification
CE marked
Operating pressure
100 PSI (690 kPa) maximum
Output
RS-232 output for all variables for selected mode, no special software required
Weight
Approx. 2.4 lbs (1.1 kg)
Inlet and outlet
1/8-inch NPT (female) threads
Accessories
RS-232 cable, stand, batteries, 1/8-inch and 1/4-inch barbed fittings