ESA Functional Tests
Tracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B, E4407B and E4408B
E7402A, E7403A, E7404A, and E7405A, Option 1DN
ESA Fu
ncti
onal
T
ests
Figure 4-19 Equipment Setup
Figure 4-20 Equipment Setup
Procedure
1. Calibrate the tracking generator by pressing
System
,
Alignments
,
Align Now
,
TG
.
Connect the RF OUT to the RF INPUT when prompted.
2. Press
System
,
Power On/Preset
,
Preset Type
(Factory),
Preset
on the analyzer, then wait
for the preset routine to finish. Set the analyzer by pressing the following keys:
FREQUENCY
,
Center Freq
, 50,
MHz
CF Step
, 100,
MHz
SPAN
,
Zero Span
AMPLITUDE
,
More
,
Y Axis Units
,
dBm
Source,
Amplitude
(On), –20,
dBm
System
,
Alignments
,
Auto Align
,
Off
3. Zero and calibrate the power meter and power sensor. Make sure the power meter is
reading out in dBm. Enter the power sensor 50 MHz cal factor into the power meter.
4. Connect the power sensor to the RF Out on the analyzer as shown in
.
5. Press REL on the power meter. The power meter readout amplitudes are now relative to
the power level at 50 MHz.
Содержание ESA Series
Страница 6: ...6 Contents Table of Contents ...
Страница 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Страница 33: ...33 Instrument Messages 1 Instrument Messages ...
Страница 111: ...111 Functional Tests 2 Functional Tests ...
Страница 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Страница 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Страница 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...