Chapter 3
121
PSA Functional Tests
Amplitude Accuracy at 50 MHz
PSA F
uncti
on
al
T
ests
Figure 3-1. Amplitude Accuracy Test Setup
Procedure
1. Zero and calibrate the power meter.
2. Configure equipment as shown in
, with the power splitter connected to the
spectrum analyzer input.
CAUTION
To minimize stress on the test equipment connections, support the power
sensor.
3. If the auto alignment for the analyzer has not been performed within the past 24 hours,
press
System
,
Alignments
,
Align All Now
to perform the auto alignment routine.
4. Press
Preset
(Factory Preset) on the analyzer.
5. Set up the spectrum analyzer by pressing:
Frequency
, 50,
MHz
Span
, 2,
MHz
Input/Output
,
RF Coupling
(DC)
Amplitude
,
Attenuation
, 10,
dB
Sweep
,
Auto Swp Time
,
Accy
6. Set up the synthesized sweeper by pressing:
CW
, 50,
MHz
Power Level
, -4,
dBm
RF
(On)