![Agilent Technologies E8363B Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/agilent-technologies/e8363b/e8363b_installation-note_2868101018.webp)
18
Installation Note E8364-90024
Analyzers with Hardware Options UNL, 014, 016, and 080
Refer to
for this portion of the procedure. The new parts referenced in this
procedure are listed in
Install the following cables (provided)
in the order listed
.
• Wire harness
8121-0966 A45 reference switch
①
to A16 test set motherboard J31
• Wire harness
8121-0966 A46 reference switch
②
to A16 test set motherboard J30
• W96 E8364-20127 A45 reference switch to A46 reference switch
• W97 E8364-20121 A46 reference switch to REFERENCE 1 SOURCE OUT
• W98 E8364-20123 A45 reference switch to REFERENCE 1 RCVR R1 IN
• W99 E8364-20125 A45 reference switch to A28 channel R1 mixer
• W95 E8364-20119 A46 reference switch to channel R1 detector
Figure 13 Cable Installation, Analyzers with Options UNL, 014, 016, and 080