5-20
Service Guide E8364-90026
Theory of Operation
PNA Series Microwave Network Analyzers
Signal Separation Group Operation
E8362B, E8363B, E8364B
High Dynamic Range Configuration
With a few jumper changes, you can configure the Option 014 for higher dynamic range
measurements. By swapping the front panel jumpers for one port, signal flow through the
corresponding coupler is reversed, increasing the test signal sensitivity by 15 dB.
In the forward direction, as shown in
, the signal flow through the test port
coupler (A26) is reversed by arranging the front panel jumpers such that RCVR B IN
connects to CPLR THRU and CPLR ARM connects to SOURCE OUT.
In the reverse direction, not shown, the signal flow through the test port coupler (A25) is
reversed by arranging the front panel jumpers such that RCVR A IN connects to
CPLR THRU and CPLR ARM connects to SOURCE OUT.
For more information on high dynamic range measurements and configurations, search for
“Option 014” in the embedded help index in the analyzer.
Figure 5-5 High Dynamic Range Configuration in the Forward Direction
Содержание E8362B
Страница 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Страница 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Страница 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Страница 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Страница 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Страница 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Страница 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...