7-60
Service Guide E8364-90001
Repair and Replacement Procedures
PNA Series Microwave Network Analyzers
Removing and Replacing the A41 Hard Disk Drive
E8362A, E8363A, E8364A
Removing and Replacing the A41 Hard Disk Drive
Certain unique files exist on the hard disk drive that are necessary for proper operation of
your analyzer. These files must be copied to another location to allow them to be installed
onto the new hard disk drive after it has been installed.
If you are replacing the hard disk, not just removing the hard disk drive assembly (HDDA)
to allow access to other assemblies, the following procedure must be performed first.
If you are not replacing the actual hard disk, proceed to
Copy Unique Files from the Hard Disk
The files to be copied include:
• license file (
gen.lic
)
• mixer calibration files (
mxcalfiles_pxx
)
• user calibration kit file (
user_calkitfile
)
• any user data files or program files that need to be preserved
The network analyzer must be powered up and operating for this procedure.
1. Insert a writable floppy disk into the floppy disk drive.
2. Open two copies of
Windows
Explorer
:
a. Click on
START
. (It may be necessary to exit the
Network Analyzer
application.)
b. Point to
Programs
,
Accessories
, and then click on
Windows
Explorer
.
c. Adjust the horizontal width of the
Windows
Explorer
window to approximately
one-half the width of the display and position it to either the left or right side of the
display.
d. Repeat this procedure to create another copy of the
Windows
Explorer
window and
position it to the opposite side of the display so that the two copies are displayed
side-by-side.
3. Set one copy of
Windows
Explorer
to the
A:
drive and the other to the
C
: drive:
a. In one copy of
Windows
Explorer
, double-click on
My
Computer
, and then
double-click on
3 1/2 Floppy (A:)
. A listing of the files on the floppy disk will appear
in the right-hand side of the window.
b. In the other copy of
Windows
Explorer
, double-click on
My
Computer
, and then
double-click on
Hard Disk (C:)
.
c. Navigate to
C:\Program Files\Agilent\Network Analyzer
by double-clicking on
each of the names in this path name, in turn. A listing of the files in this directory on
the hard disk will appear in the right-hand side of the window.
4. Copy each of the following files from the hard disk drive to the floppy disk drive:
a. Files to copy:
gen.lic
user_calkitfile
mxcalfile_pxx
(where
xx
is a 1 or 2 digit number)
Any personal user files that you wish to preserve.
Содержание E8362A
Страница 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Страница 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Страница 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Страница 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Страница 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Страница 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Страница 287: ...Service Guide E8364 90001 A 1 A Error Terms ...