Installation Note E8361-90003
3
Contacting Agilent
Assistance with test and measurements needs and information on finding a local Agilent
office are available on the Web at:
www.agilent.com/find/assist
If you do not have access to the Internet, please contact your Agilent field engineer.
NOTE
In any correspondence or telephone conversation, refer to the Agilent product by
its model number and full serial number. With this information, the Agilent
representative can determine whether your product is still within its warranty
period.
About Installing the Upgrade Kit
IMPORTANT
Option 016 can only be installed on analyzers with the option
combinations listed below as “Products affected”. If your analyzer does not
have the proper options (as listed below), it will be necessary to install
those options BEFORE installing this option (Option 016).
Description of Option 016 Receiver Attenuators
This option adds a 50-dB step attenuator in the channel A and channel B signal paths.
Products affected. . . . . . . . . . . . . . . . . . . . . . . . E8361A/C with Options 014 and UNL (with or
without Options 080 and 081)
Installation to be performed by . . . . . . . . . . . . Agilent service center or personnel qualified
by Agilent
Estimated installation time . . . . . . . . . . . . . . . 1.0 hour
Estimated adjustment time . . . . . . . . . . . . . . . 1.0 hour
Estimated full instrument calibration time . . 4.5 hours