18
Installation Note E8356-90052
Receiver Calibration Adjustment
The receiver calibration is used to adjust the network analyzer receivers for a flat response
across its full frequency range. A power meter is connected to Port 1, as shown in
, to establish a reference for flatness. A signal from Port 1, as shown in
, is used to adjust the “B” receiver at Port 2. The adjustment is repeated using a
signal from Port 2 to adjust the “A” receiver at Port 1.
1. Connect the equipment as shown in
. Connect a GPIB cable between the network
analyzer and the power meter.
Figure 8
Setup for the Receiver Calibration Adjustment
Figure 9
Setup for the Receiver Calibration Adjustment of Port 2
2. On the
System
menu, point to
Service
,
Adjustments
, then click
Receiver Calibration
.
3. Ensure the GPIB settings are correct.
4. Click
Calibrate
, then follow the instructions as they are displayed.