![Agilent Technologies E5070B Скачать руководство пользователя страница 535](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065535.webp)
Chapter 15
529
Measurement Examples
Executing Power Calibration
15
. Me
asur
eme
n
t Exa
m
ple
s
Executing Power Calibration
This section shows an example of executing power calibration using the E4418B power
meter and the E4412A power sensor.
Overview of execution procedure
In this example, power calibration is executed according to the steps shown in Table 15-11.
Table 15-11
Execution Procedure of Power Calibration
Step
Description
“1. Connecting Power Meter” on page 529
Connects the power meter to the E5070B/E5071B.
“2. Setting Address of Power Meter” on
page 530
Configures the power meter's GPIB address with the
E5070B/E5071B.
“3. Setting Stimulus Condition” on page 530
Sets conditions such as the power level and frequency.
“4. Executing Zero Adjustment and
Calibration of Power Sensor” on page 530
Executes zero adjustment and calibration of the power
sensor.
“5. Setting Calibration Data Measurement
Conditions” on page 530
Selects the port, selects the power sensor, sets the number
of measurements at one point, and sets the
tolerance during
power calibration
.
“6. Connecting Power Sensor” on page 531
Connects the power sensor.
“7. Measuring Calibration Data” on page 531
Executes the measurement of calibration data.
1. Connecting Power Meter
Connect the E4418B (GPIB address: 14) to the E5070B/E5071B as shown in Figure 15-39.
Figure 15-39
Connecting power meter
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...