Chapter 15
523
Measurement Examples
Evaluating Transmission Characteristics of a Front End Module
15
. Me
asur
eme
n
t Exa
m
ple
s
Step 3.
Select the test ports assigned to ports 1 to 4 for channel 1.
Setting Description
Key Operation
Assign test port A to port 1.
Port1
-
A
Assign test port T1 to port 2.
Port2
-
T1
Assign test port R1+ to port 3.
Port3
-
R1+
Assign test port R1- to port 4.
Port4
-
R1-
Step 4.
Assign test ports for channels 2, 3, and 4. Press the
key to switch the active
channel and then make the setting in the same way as Step 3.
Step 5.
Enable the control of the E5091A.
Setting Description
Key Operation
Enable the control of the E5091A.
Control
(set it to
ON
)
4. Setting Control Line
Step 1.
Set the bits of the control line for channel 1.
Setting Description
Key Operation
Set channel 1 to the active channel.
Line 1: Low
Control Lines
(check only)
Line 2: High
Control Lines
-
Line 1
(set it to
High
)
Step 2.
Set the control line for channels 2, 3, and 4 according to the same procedure as in Step 1.
5. Setting Sweep Conditions
Step 1.
Set the sweep conditions for channel 1.
Setting Description
Key Operation
Set channel 1 to the active channel.
Start frequency: 400 MHz
Stop frequency: 1.4 GHz
Number of points: 201
-
Point
-
Step 2.
Set sweep conditions for channels 2, 3, and 4 according to the same procedure as in Step 1.
6. Setting Balance Conversion Topology
For channel 2, set the balance conversion topology in order to perform measurement that
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...