Chapter 11
375
Limit Test
Limit Test
11
. Lim
it T
es
t
Displaying judgment result of limit test
Judgment result of measurement points and trace
Measurement points that fail are displayed in red on the screen. The judgment result of the
trace is indicated by Pass or Fail displayed in the upper right part of the graph.
Figure 11-1
Result display of measurement points and trace
Judgment result of channels
If a channel has a judgment result of fail, the message shown in Figure 11-2 appears on the
screen
(it will be judged as failed if one or more unsatisfactory trace exists in any of the
limit test, the ripple test or the bandwidth test within the channel.)
Figure 11-2
Channel fail message
Follow these steps to turn on/off the display of the channel fail message.
Step 1.
Press
.
Step 2.
Press
Limit Test
.
Step 3.
Press
Fail Sign
. Each press toggles between on/off.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...