![Agilent Technologies E5070B Скачать руководство пользователя страница 299](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065299.webp)
Chapter 7
293
Fixture Simulator
Evaluating Balanced Devices (balance-unbalance conversion function)
7
. F
ixt
ur
e
Sim
ula
to
r
Evaluating Balanced Devices (balance-unbalance conversion
function)
The balance-unbalance conversion function simulates a measurement under a balanced
state based on measurement results obtained in an unbalanced state (Figure 7-7). This
function enables you to evaluate devices with balanced ports.
NOTE
Be sure to set the impedances of the two unbalanced ports equal to each other. For more
details on setting up port impedance for unbalanced ports, refer to “Converting the Port
Impedance of the Measurement Result” on page 286.
Figure 7-7
Balance-unbalance conversion
The types of devices that can be evaluated using the E5070B/E5071B are shown in Figure
7-8.
NOTE
To evaluate a balanced device, an E5070B/E5071B with at least three test ports (option
313, 314, 413, or 414) is required.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...