134
Chapter 4
Calibration
Full 3-Port and Full 4-Port Calibration using 2-Port ECal
Figure 4-27
ECalAssistant (Connection) dialog box
Step 14.
Re-connect the ECal module following the instructions given in each dialog box and
continue the calibration process.
Step 15.
When all calibration data have been collected, a dialog box with the
Complete!
sign
appears as shown in Figure 4-28. Press the
Done
button to finish the calibration. If you
wish to cancel the calibration, press the
Cancel
button.
Figure 4-28
EcalAssistant (Finish) dialog box