Chapter 4
113
Calibration
1-Port Calibration (reflection test)
4. Cal
ibra
tion
1-Port Calibration (reflection test)
In 1-port calibration, calibration data are measured by connecting an OPEN standard, a
SHORT standard, and a LOAD standard to the desired test port. This calibration effectively
eliminates the frequency response reflection tracking error, directivity error, and source
match error from the test setup in a reflection test using that port (Figure 4-11).
Figure 4-11
1-Port error model (1-port calibration)
Figure 4-12
Connecting the standard for 1-port calibration
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...