58
Chapter 3
Quick Start Guide
Measurement Example of a Bandpass Filter
STEP 4. Analyzing Measurement Results
This section describes how to use the marker function to read out important parameters for
the transmission measurement of the bandpassfilter (insertion loss, -3 dB bandwidth).
Measuring the Insertion Loss
Step 1.
Display a marker.
•
[Marker] - Marker 1
Step 2.
Using one of the following methods, move the marker to the center frequency of the
bandpass filter.
•
On the entry bar, press
[9] [4] [7] [.] [5] [M/
μ
]
.
•
Turn the rotary knob (
) on the front panel to set it to the center frequency (947.5
MHz).
Step 3.
Read the marker value displayed as shown in Figure 3-17. In this example, the response
value denotes the insertion loss.
Figure 3-17
Measuring an Insertion Loss