7
371
SCPI Command Reference
:TRIGger:FP[1-1]:MODE
7. SCPI Command
Re
fe
re
nc
e
Parameter
Equivalent key
PN Menu -> Trigger -> Ext Trig Polarity
SP Menu -> Trigger -> Ext Trig Polarity
FP Menu -> Trigger -> Ext Trig Polarity
TR Menu -> Trigger -> Ext Trig Polarity
:TRIGger:FP[1-1]:MODE
Syntax
*1
:TRIGger:FP[1-1]:MODE {ANALyzer|TESTer}
:TRIGger:FP[1-1]:MODE?
Description
Selects measurement mode in the Frequency. power and DC currnet measurement mode
Parameter
Equivalent key
FP Menu -> Trigger -> Mode
:TRIGger:FP[1-1]:SOURce
Syntax
:TRIGger:FP[1-1]:SOURce {INTernal|EXTernal|MANual|BUS}
:TRIGger:FP[1-1]:SOURce?
Description
Selects trigger source
Parameter
Description
NEGative(Preset
value)
Set External trigger polarity to 'NEGative'
POSitive
Set External trigger polarity to 'POSitive'
*1.“Option not installed” error message is generated when setting the trigger mode to the analyzer mode with the
option 011 instrument.
Description
ANALyzer(Preset
value)
Set trigger mode to 'ANALyzer'
TESTer
Set trigger mode to 'TESTer'
Description
INTernal(Preset
value)
Set trigger source to 'INTernal'
Содержание E5052A
Страница 4: ...4 ...
Страница 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Страница 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Страница 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Страница 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Страница 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...
Страница 476: ...476 AppendixB Status Reporting System Status Register Structure Figure B 3 Status register structure 1 of 6 ...
Страница 478: ...478 AppendixB Status Reporting System Status Register Structure Figure B 5 Status register structure 3 of 6 ...
Страница 480: ...480 AppendixB Status Reporting System Status Register Structure Figure B 7 Status register structure 5 of 6 ...