2
Power Meter Operation
70
E4418B Power Meter User’s Guide
Setting Measurement Limits
You can configure the power meter to detect when a measurement is
outwith a predefined upper and/or lower limit value.
There are two types of measurement limits you can set:
•
Channel Limits - are applied to the input channel and are for power
measurements only.
•
Window Limits - are windows based (upper and lower) and can be
applied to power, ratio, or difference measurements. In addition, the
window based limits can be set to output a TTL logic level at the rear
panel Rmt I/O port when the predefined limits are exceeded.
Setting Channel Limits
The power meter can be configured to verify the power being measured
against an upper and/or lower limit value. The range of values that can be
set for upper and lower limits is –150 dBm to 230 dBm. The default
upper limit is 90.00 dBm and the default lower limit is –90.00 dBm.
To set the limits:
1
Press
,
,
. The current setting of the
maximum and minimum limits are displayed under the
and
softkeys respectively.
2
To change either of these settings, press the appropriate softkey. The
power meter displays the current value in a pop- up window. Modify
this value (see below) until the desired value is displayed.
•
Use
or
to modify the digit on which the cursor is
currently positioned.
•
Use
or
to move to other digits.
3
To confirm your choice, press the appropriate measurement units.
Limits can be disabled and re- enabled simply by pressing
.
N O T E
Only one set of limits can be on at a time, that is, Channel, OR Window.
Input Settings
Limits
Max
Min
Limits Off On
Содержание E4418B
Страница 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Страница 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 47: ...Power Meter Operation 2 E4418B Power Meter User s Guide 23 Contacting Agilent Technologies on page 103 ...
Страница 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...
Страница 126: ...2 Power Meter Operation 102 E4418B Power Meter User s Guide Figure 2 23 Replacing the fuse In line fuse Spare fuse ...
Страница 168: ...3 Menu Reference 144 E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 169: ...145 E4418B Power Meter User s Guide Agilent Technologies 4 Error Messages Introduction 146 Error Messages 148 ...