Power Meter Operation
2
E4418B Power Meter User’s Guide
35
Zeroing and Calibrating Using TTL Inputs
You can use the TTL inputs on the rear panel Rmt I/O port to initiate
zero and calibration cycles on the power meter. The connector is an RJ- 45
Series shielded modular jack with the TTL input pins connected as shown
in
.
Figure 2-3
Rmt I/O Port TTL Inputs
The TTL inputs are active low and control the zero and calibration functions
as shown in
Pin Number
Connection
1
none
2
Ground
3
Upper Window TTL Output
4
Lower Window TTL Output
5
TTL Input 1
6
TTL Input 2
7
Ground
8
Ground
1 2 3 4 5 6 7 8
Table 2-2
TTL Inputs Control Logic
Input 1
Input 2
Operation
1
1
None
1
0
CAL
0
1
ZERO
0
0
CAL
Содержание E4418B
Страница 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Страница 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 47: ...Power Meter Operation 2 E4418B Power Meter User s Guide 23 Contacting Agilent Technologies on page 103 ...
Страница 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...
Страница 126: ...2 Power Meter Operation 102 E4418B Power Meter User s Guide Figure 2 23 Replacing the fuse In line fuse Spare fuse ...
Страница 168: ...3 Menu Reference 144 E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 169: ...145 E4418B Power Meter User s Guide Agilent Technologies 4 Error Messages Introduction 146 Error Messages 148 ...