Hub Downstream Signal Quality Testing
5
DSOX4USBSQ USB 2.0 Signal Quality Analysis Application Electrical Testing Notes
77
7
Press the
Auto Setup
softkey to automatically set up the oscilloscope for
the selected test.
For optimal test results, you can make adjustments to the auto setup.
Generally, you want one packet of data, plus the time of one bit on
each end, across the oscilloscope's display. Also, vertical scale and
offset should be adjusted so that the signals take six vertical divisions
without being clipped.
You should see the transmitted test packet on the oscilloscope as below.
8
Press the
Run Test
softkey to run the selected signal quality test.
When tests are run, the oscilloscope stops acquisitions if they are
running, analyzes the data on screen, and then displays the results.
When the test completes, the results are automatically displayed. You
can press the
Display Results
softkey to disable or reenable the test
results display.
The analyzed acquisition remains on screen and can be viewed by
moving or dismissing the results dialog box.
N O T E
Tests can take several minutes to run and cannot be cancelled once started.