410
Agilent 6000 Series Oscilloscopes Programmer's Reference
3
Commands by Subsystem
:TRIGger:SEQuence:EDGE
(see
Command Syntax
:TRIGger:SEQuence:EDGE{1 | 2} <source>, <slope>
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital0,..,DIGital15} for the MSO models
<slope> ::= {POSitive | NEGative}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger:SEQuence:EDGE<n> command defines the specified sequencer
edge resource according to the specified <source> and <slope>. To disable
an edge resource, set its <source> to NONE. In this case, <slope> has no
meaning.
Query Syntax
:TRIGger:SEQuence:EDGE{1 | 2}?
The :TRIGger:SEQuence:EDGE<n>? query returns the specified sequencer
edge resource setting. If the edge resource is disabled, the returned
<source> value is NONE. In this case, the <slope> is undefined.
Return Format
<source>, <slope><NL>
See Also
•
"Introduction to :TRIGger Commands"
•
•
•
•
Содержание DSO6014A/L
Страница 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Страница 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Страница 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Страница 584: ...584 Agilent 6000 Series Oscilloscopes Programmer s Reference 8 More About Oscilloscope Commands...
Страница 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...