Specifications
16
Agilent Cary 60 Spectrophotometer User’s Guide
Environmental Conditions
No draughts
No corrosive fumes
No vibration
NOTE
For
optimum analytical performance
, it is recommended that the ambient
temperature of the laboratory be between
20 and 25 °C
and be held constant to
within
±
2 °C throughout the entire working day.
The area should have a dust-free, low humidity atmosphere. Air
conditioning is recommended. The room should be temperature-
controlled if your analyses are particularly sensitive.
Condition
Altitude (m, ft)
Temp. (°C, °F)
Humidity (%RH)
non-condensing
Non-operating (transport)
0–4600, 0–15000
-40–70, -40–158
<90
Operating within performance
specifications
0–3100, 0–10000
5–40, 41–104
15–80