9
1 Safety guidelines
Safety equipment and features
BenchCel Microplate Handler User Guide
WARNING
To reduce the risk of injury from moving parts, ensure that the BenchCel shield and
the shields for any integrated devices are installed on the workstation.
WARNING
Operating the Bravo Platform or Microplate Centrifuge with Loader without a safety
shield or a light curtain connected to the safety interlock circuit increases the risk of
injury.
Safety interlock circuit
WARNING
To avoid potential injury, Agilent recommends that you install a main emergency
stop button (Robot Disable Hub) that will stop the BenchCel robot and all other
devices on the safety interlock circuit simultaneously. In addition, all operators must
be instructed in the emergency stop procedure.
The BenchCel Microplate Handler has a safety interlock circuit that is designed to
protect operators from moving-parts hazards. The safety interlock circuit must be
closed for the device to operate. The pendant connects to the safety interlock circuit.
Pressing the red button on the pendant trips the interlock circuit and causes the motion
of the BenchCel operation to stop.
Figure
Emergency-stop pendant
When the BenchCel device is integrated with another device that has a safety interlock
circuit, such as the Bravo Platform, Agilent Technologies uses the Robot Disable Hub to
connect all the devices on the interlock circuit. Pressing the red button on a pendant or
interrupting a light curtain connected to the Robot Disable Hub will trip the safety
interlock circuit and stop all the devices connected to the circuit.
Figure
Robot Disable Hub