![Agilent Technologies B1500A Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/agilent-technologies/b1500a/b1500a_user-manual_2867697032.webp)
1-10
Agilent B1500A User’s Guide, Edition 10
Getting Started
To Use Tracer Test Mode
To Use Tracer Test Mode
The following procedure performs the tracer test setup.
1. Click the Tracer Test tab on the main screen. The tab is in the leftmost column
on the screen. The I/V Trace screen is displayed. See Figure 1-4.
2. Set the channels at the field below the graph as shown in Table 1-1.
3. Set the measurement condition at the right field of the graph as shown in Table
1-3.
4. Set the graph display condition at the field below the graph as shown in Table
1-2.
To enter the value, use the USB keyboard or the screen/numeric keyboard opened by
clicking the right button of the entry field.
To start measurement, click
for the single measurement or
for the repeat measurement.
If the Stop field is highlighted, the range of measurement and graph display can be
changed by using the rotary knob during the repeat measurement.
Table 1-1
I/V Trace Channel Setup
Table 1-2
I/V Trace Graph Setup
Unit
V Name
I Name
Mode
a
a. Channel output mode. Set
I
or
IPULSE
for the current output and volt-
age measurement. Set
V
or
VPULSE
for the voltage output and current
measurement.
Function
SMU1
V1
I1
V
VAR1
SMU2
V2
I2
V
CONST
Name
Scale
Name
Scale
Y
I1
LIN
X
V1
LIN
Содержание B1500A
Страница 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Страница 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Страница 23: ...1 Getting Started ...
Страница 41: ...2 Introduction ...
Страница 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Страница 75: ...3 Installation ...
Страница 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Страница 133: ...4 Specifications ...
Страница 179: ...5 Accessories and Options ...