![Agilent Technologies B1500A Скачать руководство пользователя страница 113](http://html.mh-extra.com/html/agilent-technologies/b1500a/b1500a_user-manual_2867697113.webp)
Agilent B1500A User’s Guide, Edition 10
3- 39
Installation
Mounting Connectors
To Connect MFCMU Output
The MFCMU output connection example is shown in Figure 3-17. This connection
is for the MFCMU without the SCUU. Then Agilent N1300A-001/002 CMU cable
should be used for the connection between the MFCMU and the connectors you
mount. It is ok to leave open the ground wire extended from the CMU cable.
Figure 3-17
MFCMU Output Connection
The Hcur/Hpot/Lcur/Lpot lines should be connected to the DUT as shown above.
Use a 50
coaxial cable (Agilent part number: 8120-0367) for the
Hcur/Hpot/Lcur/Lpot lines connection. Use a wire for the guard connection.
+FXU
*XDUG
+SRW
/SRW
*XDUG
/FXU
,QVXODWRU
3ODWH
%1&FRQQHFWRU
&RD[LDOFDEOH
:LUH
WR'87
KLJKWHUPLQDO
WR'87
ORZWHUPLQDO
Содержание B1500A
Страница 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Страница 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Страница 23: ...1 Getting Started ...
Страница 41: ...2 Introduction ...
Страница 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Страница 75: ...3 Installation ...
Страница 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Страница 133: ...4 Specifications ...
Страница 179: ...5 Accessories and Options ...