Agilent B1500A User’s Guide, Edition 10
3- 33
Installation
Mounting Connectors
To Connect SMU Output
The SMU output connection examples are shown in Table 3-8.
WARNING
Potentially hazardous voltages, up to ± 100 V (MPSMU/HRSMU) or ± 200 V
(HPSMU), are present at the Force, Sense, and Guard terminals.
To prevent electrical shock, do not expose these lines.
Before turning Agilent B1500A on, connect the Interlock terminal to an
interlock circuit.
Before you touch any connections to these terminals, turn Agilent B1500A off,
disconnect the power cable, and discharge any capacitors.
CAUTION
Never connect the Guard terminal to any output, including circuit common, chassis
ground, or any other guard terminal. Doing so will damage the SMU.
NOTE
Low-Noise Coaxial Cable
For the extended measurement paths over the connector plate, use low-noise coaxial
cable (Agilent part number 8121-1191). This cable can maximize the guard effects
and minimize the impression of the external noise.
Figure 3-12 shows the cutting example of this cable. Key point is the isolation
between the conductive layer and the center conductor. So, cut and trim the end of
the cable as shown in this figure by using a cutter and so on.
Figure 3-12
Coaxial Cable Cutting Example
,QVXODWRUEODFN
2XWHUFRQGXFWRU
IRU*XDUGVLJQDO
,QVXODWRUFOHDU
&RQGXFWLYHOD\HUEODFN
&HQWHUFRQGXFWRU
IRU)RUFH6HQVHVLJQDO
DSSUR[WRPP
PLQPP
&RYHUKHUHXVLQJVOHHYH
Содержание B1500A
Страница 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Страница 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Страница 23: ...1 Getting Started ...
Страница 41: ...2 Introduction ...
Страница 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Страница 75: ...3 Installation ...
Страница 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Страница 133: ...4 Specifications ...
Страница 179: ...5 Accessories and Options ...