36
G1369B LAN Interface User Manual
2
Getting Started
Storing the Settings Permanently with Bootp Program
Storing the Settings Permanently with Bootp Program
If you want to change parameters of the card using the Bootp follow the
instructions below.
1
Turn off the module that hosts the LAN Interface and remove the card.
2
Change the card’s settings of the Configuration Switch to
”Bootp & Store”
mode, see
Table 5
on page 22.
3
Install the LAN Interface.
4
Start the CAG Bootp Server program and open its window.
5
If required, modify the parameters for the LAN Interface according to your
needs using the existing configuration.
6
Press
OK
to exit the Bootp Manager.
7
Now turn on the module with the LAN Interface and view the Bootp Server
window. After some time the Bootp Server will display the request from the
LAN Interface. The parameters are now stored permanently in the
non-volatile memory of the card.
8
Close the CAG Bootp Server program and turn off the module and remove
the LAN Interface.
9
Change the settings of the card’s Configuration Switch to
“Using Stored”
mode, see
Table 5
on page 22.
10
Install the card and power cycle the module with the LAN Interface. The
card can be accessed now via LAN without the CAG Bootp Server program,
refer to
“PC and Agilent ChemStation Setup”
on page 48.
N O T E
Use an ESD (Electro-Static Discharge) wrist strap when handling electronics. Refer to your
instrument manual for details.