STEP
Dionysos
Job: 210%XXXX5511-0002
5965-5601E
Comp: 04-28-00
Time: 11:20
OP: NN
PDFINFO H 5 6 0 1 - 0 2
2
Product Specifications
Input
DC:
Maximum voltage (any terminal to any other terminal or
chassis):
120 Vdc
AC rms:
Maximum voltage (any terminal to any other terminal or
chassis):
120 V rms
Maximum current (per channel common, non-
inductive):
50 mA
Maximum power per channel:
1 VA
DC
Maximum thermal offset
per channel, differential
Hi-Lo:
4
µ
V
Closed channel resistance:
100
Ω
± 10%
Insulation resistance
(between any two points):
10E9
Ω
Insulation resistance
(Hi to Lo, power off):
n/a
AC
Minimum bandwidth
(-3 dB, 50
Ω
source/load):
10 MHz (protection resistors shorted)
Crosstalk (channel-to-
channel):
100 kHz:
–70 dB
10 MHz:
–20 dB
Both:
n/a
Closed channel
capacitance:
<
150 pF Hi-Lo,
<
150 pF Lo-Guard,
<
2000 pF
Guard-Chassis
General Characteristics
Relays:
Reed relays
Break-before-make
Relays open on power down
Relays open on power up
Minimum relay life:
No load:
10E8 operations
Rated load:
10E7 operations
Screw terminal wire size:
16 to 26 AWG (1.5, 1.2, 0.9, 0.75, 0.5 mm)
Scanning rate:
600 channels/s typ.
General Specifications
VXI Characteristics
VXI device type:
Register based, A16, slave only
Size:
B
Slots:
1
Connectors:
P1
Shared memory:
None
VXI busses:
None
C-size compatibility:
Requires E1403C
Instrument Drivers
See the Agilent Technologies Website (http://www.agilent.com/find/
inst_drivers) for driver availability and downloading.
Command module
firmware:
Downloadable
Command module
firmware rev:
A.01
I-SCPI Win 3.1:
Yes
I-SCPI Series 700:
Yes
C-SCPI LynxOS:
Yes
C-SCPI Series 700:
Yes
Panel Drivers:
Yes
VXI
plug&play Win
Framework:
Yes
plug&play Win 95/NT
Framework:
Yes
VXI
plug&play HP-UX
Framework:
No