How to Measure Power
Power Measurement
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
89
5
Press
Enter
.
You can generate input triggers in any of the following ways:
•
trigger the Input Trigger Connector on the rear panel of your instrument,
•
set <
Trigger Configuration
> to <
Loopback
> so that an output trigger
automatically generates an input trigger, or
•
use the
:TRIGger
GPIB command, see your mainframe’s Programming Guide.
How to Use Output Triggering
You can configure Agilent 8163A Series Power Meter modules to output a trigger
when the instrument performs certain tasks.
To set your module’s Output Trigger Configuration:
1
See
“How to Set the Trigger Configuration” on page 62
for how to configure
the trigger connectors.
2
Move to the Power Meter channel and press the [Menu] softkey.
3
Move to <
Output Trigger Mode
>, by using the cursor key, and press
Enter
.
You will see the screen in Figure 59.
4
Move to one of the following, by using the cursor key:
–
<
Averaging Time Over
>, a trigger is output when the averaging time period
of a measurement finishes.
–
<
Meas Started
>, a trigger is output when the averaging time period of a
measurement begins.
–
<Disabled
>, the output trigger mode is disabled.
5
Press
Enter
.
Figure 59
Output Trigger Mode