Controlling the Sampler
4
Installation, Operation, and Maintenance
87
Part 2, Operation
Setting Injector Parameters on a 6890 GC
Press
[Front Injector]
or
[Back Injector]
.
Scroll to the desired setpoint detailed below.
Enter a setpoint value, or turn the setpoint on or off.
Injection volume
—
Sample volume to be injected. On a 6890N GC press
[Mode/Type]
to select. For a 6890N GC, the available volumes depend on the
syringe size configured. See
“Configuring the injector on a 6890 GC"
•
The selections represent 2%, 10%, 20%, 30%, 40%, and 50% of syringe size
•
Turn the injection volume Off to disable the injector
On a 6890A or Plus GC enter a value from the volume unit column in
FRONT INJECTOR
Injection vol
1.0<
# Sample pumps
1
Viscosity delay
1
# Sample washes
1
# Solv A washes
1
# Solv B washes
0
Slow plunger
Off
Pre dwell time
0.10
Post dwell
0.10
Sampling offset
2.0
# Solv A pre wash
2
# Solv B pre wash
2
Содержание agilent 7683B
Страница 1: ...Agilent Technologies Agilent 7683B Automatic Liquid Sampler Installation Operation and Maintenance...
Страница 10: ...10 Installation Operation and Maintenance...
Страница 18: ...18 Installation Operation and Maintenance...
Страница 70: ...70 Installation Operation and Maintenance Part 1 Installation 2 Installation...
Страница 72: ...72 Installation Operation and Maintenance...
Страница 128: ...128 Installation Operation and Maintenance Part 2 Operation 6 Vials and Bottles...
Страница 166: ...166 Installation Operation and Maintenance Part 3 Maintenance and Troubleshooting 9 Faults and Errors...
Страница 202: ...202 Installation Operation and Maintenance A Cables and Connectors...