Installation
2
Installation, Operation, and Maintenance
67
Part 1, Installation
•
If you are using the G2912A ALS Controller, a copy of this utility is included
on the G2911AA software CD. This copy includes the minimum firmware
revisions required for 7683B compatibility. For best results, check the Web
site for the latest revisions.
To check and update firmware:
1
Install the firmware update utility on a computer that is either already
connected to the devices (such as a computer running an Agilent data
system) or one that can be easily connected by a cable.
2
Connect to the first device. The utility’s help system describes in detail how
to do this. In general, use an existing LAN connection (if available), or
connect directly using an RS-232 cable between the computer and the
device.
3
Use the utility to display the firmware revisions of connected hardware.
4
If any firmware does not meet the requirements listed in
, update
it.
If a device meets the minimum requirements but a newer version is
available, Agilent recommends updating it anyway to take advantage of the
latest improvements.
6890A, 6890 Plus and 5890 Series II GCs
The utility can display the firmware revision for these GCs but cannot update
them. These GCs use programmable chips that must be physically replaced.
Contact your local Agilent representative.
Other ways to check firmware revisions
All GCs
Power cycle the instrument. When it reboots, it displays its firmware
revision.
All 6890 GCs
Press
Options
, then select
Diagnostics/Instrument Status
. Scroll
down to see GC and injector firmware revisions and dates. The G2612A card
used in the 6890 Plus is also displayed. For 6890N the tray firmware is also
displayed.
6850 GCs
Use the handheld controller. From
Status
, select
Service/Update
to
display GC and injector information.
Содержание agilent 7683B
Страница 1: ...Agilent Technologies Agilent 7683B Automatic Liquid Sampler Installation Operation and Maintenance...
Страница 10: ...10 Installation Operation and Maintenance...
Страница 18: ...18 Installation Operation and Maintenance...
Страница 70: ...70 Installation Operation and Maintenance Part 1 Installation 2 Installation...
Страница 72: ...72 Installation Operation and Maintenance...
Страница 128: ...128 Installation Operation and Maintenance Part 2 Operation 6 Vials and Bottles...
Страница 166: ...166 Installation Operation and Maintenance Part 3 Maintenance and Troubleshooting 9 Faults and Errors...
Страница 202: ...202 Installation Operation and Maintenance A Cables and Connectors...