Installation
2
Installation, Operation, and Maintenance
51
Part 1, Installation
Procedure 6. Installing the G2614A Tray
1
Remove the three inlet cover Torx screws (
).
2
Align the tray over the two left holes in the inlet cover. Tighten the short
screw in the center of the tray mounting arm to hold the tray. Install the
two long mounting screws into the tray and tighten until the tray is secure.
3
Install the tray quadrants (
). Be sure to match the numbers on the
quadrants with the tray base numbers.
Slide the leading edge of each quadrant under the guides on the tray, then
snap the front tab into the slot on the tray.
Figure 21
Installing the tray
Tray
Screw (short)
Remove screws from cover
Inlet cover
Tray
mounting
screws
(long)
C A U T I O N
Do not move or manipulate the robotic arm or gripper. Moving the arm or gripper back
and forth can cause damage. If necessary, rotate the arm assembly a few degrees until
there is enough clearance.
Содержание agilent 7683B
Страница 1: ...Agilent Technologies Agilent 7683B Automatic Liquid Sampler Installation Operation and Maintenance...
Страница 10: ...10 Installation Operation and Maintenance...
Страница 18: ...18 Installation Operation and Maintenance...
Страница 70: ...70 Installation Operation and Maintenance Part 1 Installation 2 Installation...
Страница 72: ...72 Installation Operation and Maintenance...
Страница 128: ...128 Installation Operation and Maintenance Part 2 Operation 6 Vials and Bottles...
Страница 166: ...166 Installation Operation and Maintenance Part 3 Maintenance and Troubleshooting 9 Faults and Errors...
Страница 202: ...202 Installation Operation and Maintenance A Cables and Connectors...