Installation, Operation, and Maintenance
Part 3, Maintenance and Troubleshooting
8
Maintenance
9
When the injector stops and all four indicator lights flash, install the
calibration probe into the syringe carrier. The probe installs just like a
syringe (
10
Press the Alignment button again.
a
The syringe carriage moves down and the turret rotates to verify that the
probe is installed.
b
The turret rotates the calibration hole under the probe.
c
The probe moves down to calibrate the turret to the syringe carriage.
11
When alignment is complete, the orange Align Mode light will go out.
Remove the calibration probe and secure it in its clamp below the turret
using the screw removed in
. See
If the Align Mode light stays on after recycling power, see
If this alignment was user-initiated (orange Align Mode light was off before
pressing the Alignment button) and alignment fails, unplug the injector,
reconnect it, and it will reset to the old alignment values.
12
Install the needle support assembly and syringe.
Figure 63
Installing the calibration probe
Calibration probe
Содержание agilent 7683B
Страница 1: ...Agilent Technologies Agilent 7683B Automatic Liquid Sampler Installation Operation and Maintenance...
Страница 10: ...10 Installation Operation and Maintenance...
Страница 18: ...18 Installation Operation and Maintenance...
Страница 70: ...70 Installation Operation and Maintenance Part 1 Installation 2 Installation...
Страница 72: ...72 Installation Operation and Maintenance...
Страница 128: ...128 Installation Operation and Maintenance Part 2 Operation 6 Vials and Bottles...
Страница 166: ...166 Installation Operation and Maintenance Part 3 Maintenance and Troubleshooting 9 Faults and Errors...
Страница 202: ...202 Installation Operation and Maintenance A Cables and Connectors...