34980A Service Guide
5
Disassembly and Repair
KOM Removal
1
Remove the power supply (see the procedure on
2
Disconnect the two ribbon cables on the top of the chassis.
3
Use a 3/16” nut driver to remove the nuts holding the Ext Trig DB9 connector on the
rear panel. Use a 9/32” nut driver to remove the nuts holding the GP-IB connector on
the rear panel.
4
Use a T20 Torx to remove the four screws holding the KOM printed circuit assembly
to the mainframe. Lift out the assembly.