VOLTage Subsystem Introduction
conflict" error.
l
The high limit sets the highest output voltage allowed to be set, including DC Offset and peak amplitude. It is set in
reference to the current
setting. If the specified LOAD impedance is not present at the instru-
ment's output, then the output limit may not represent the actual voltages at the output connector. For example,
if the output impedance is set to 50 Ω, but the actual load is high impedance, then the actual output peak voltage
may be up to twice the specified limit voltage.
l
Specifying Voltage Units:
You can set the output limit voltage only in volts.
l
When
is ON, and
is ON, voltage limit settings on both channels
affect maximum amplitude and offset voltage settings on both channels. The most restrictive combination of high
and low limits from either channel is used.
[SOURce[1|2]:]VOLTage:LIMit:STATe {ON|1|OFF|0}
[SOURce[1|2]:]VOLTage:LIMit:STATe?
Enables or disables output amplitude voltage limits.
Parameter
Typical Return
{ON|1|OFF|0}, default OFF
0 (OFF) or 1 (ON)
Set and enable ±2.5 V output limits on channel 1:
VOLT:LIM:HIGH 2.5
VOLT:LIM:LOW -2.5
VOLT:LIM:STAT ON
l
When this is turned ON, if the present settings of amplitude and offset exceed the limits, then the limits will be dis-
abled. The instrument will generate either a "Settings conflict" error.
l
When
is ON, and
is ON, voltage limit settings on both channels
affect maximum amplitude and offset voltage settings on both channels. The most restrictive combination of high
and low limits from either channel is used.
l
Limits are set in reference to the current setting of
. If the specified LOAD impedance is not
present at the instrument's output, then the output limit may not represent the actual voltages at the output con-
nector. For example, if the output impedance is set to 50 Ω, but the actual load is high impedance, then the actual
output peak voltage may be up to twice the specified limit voltage.
[SOURce[1|2]:]VOLTage:OFFSet {<
offset
>|MINimum|MAXimum}
[SOURce[1|2]:]VOLTage:OFFSet? [{MINimum|MAXimum}]
Sets DC offset voltage.
Parameter
Typical Return
± 5 VDC into 50 Ω, default 0
+1.0000000000000E-01
Set offset voltage to 100 mV:
VOLT:OFFS 100 mV
l
The relationship between offset voltage and output amplitude is shown below.
324
Agilent 33500 Series Operating and Service Guide
Содержание Agilent 33500 Series
Страница 1: ...Agilent Technologies 33500 Series Waveform Generator Operating and Service Guide ...
Страница 68: ...Output Configuration OUTPUT SYNC SOURCE CH1 CH2 Agilent 33500 Series Operating and Service Guide 67 ...
Страница 410: ...Block Diagram Block Diagram Agilent 33500 Series Operating and Service Guide 409 ...
Страница 429: ......