SWEep Subsystem Introduction
Parameter
Typical Return
Enable sweep:
SWE:STAT ON
[SOURce[1|2]:]SWEep:TIME {<
seconds
>|MINimum|MAXimum}
[SOURce[1|2]:]SWEep:TIME? [{MINimum|MAXimum}]
Sets time (seconds) to sweep from start frequency to stop frequency.
Parameter
Typical Return
1 ms to 250,000 s for linear sweep, up to 500 s for logarithmic sweep;
default 1 s
+2.50000000001
Set sweep time to 25 s:
SWE:TIME 25
l
The number of discrete frequency points in the sweep is calculated based on the sweep time.
298
Agilent 33500 Series Operating and Service Guide