![Agilent Technologies Agilent 33500 Series Скачать руководство пользователя страница 248](http://html.mh-extra.com/html/agilent-technologies/agilent-33500-series/agilent-33500-series_operating-and-service-manual_2867488248.webp)
MEMory Subsystem
l
Default names are "AUTO_RECALL", "STATE_1", "STATE_2", "STATE_3", and "STATE_4".
l
You can name location 0, but the name is overwritten when power is cycled and a new power-down state is stored
there.
MEMory:STATe:DELete {0|1|2|3|4|}
Deletes a state storage location.
Parameter
Typical Return
{0|1|2|3|4|}
(none)
Delete the contents of storage location 1:
MEM:STAT:DEL 1
l
Default names are "AUTO_RECALL", "STATE_1", "STATE_2", "STATE_3", and "STATE_4".
l
Although you may delete the state in location 0, it will be restored to its
, and state name
("AUTO_RECALL") at the next power up.
l
Attempting to recall a state from an empty location generates an error.
MEMory:STATe:NAME {0|1|2|3|4} [,<
name
>]
MEMory:STATe:NAME? {0|1|2|3|4}
Names a storage location.
Parameter
Typical Return
An unquoted string of up to 12 characters.
The first character must be a letter (A-Z).
Others can be letters, numbers (0-9), or underscores ("_").
If name omitted, factory default name is used.
TEST_RACK_1
Rename location 1:
MEM:STAT:NAME 1,TEST_RACK_1
l
Default names are "AUTO_RECALL", "STATE_1", "STATE_2", "STATE_3", and "STATE_4".
l
You can name location 0, but the name is overwritten when power is cycled and a new power-down state is stored
there.
l
May assign same name to different locations.
l
Deleting a storage location's contents (
) resets associated name to factory default ("AUTO_
RECALL", "STATE_1", "STATE_2", "STATE_3", or "STATE_4").
l
State names are unaffected by
MEMory:STATe:RECall:AUTO {ON|1|OFF|0}
MEMory:STATe:RECall:AUTO?
Disables or enables automatic recall of instrument state in storage location "0" at power on.
Agilent 33500 Series Operating and Service Guide
247
Содержание Agilent 33500 Series
Страница 1: ...Agilent Technologies 33500 Series Waveform Generator Operating and Service Guide ...
Страница 68: ...Output Configuration OUTPUT SYNC SOURCE CH1 CH2 Agilent 33500 Series Operating and Service Guide 67 ...
Страница 410: ...Block Diagram Block Diagram Agilent 33500 Series Operating and Service Guide 409 ...
Страница 429: ......