Optimization of the Agilent 1260 Infinity Binary LC
How to Configure the Optimum Delay Volume
Delay Volumes in the Agilent 1260 Infinity Binary LC
on page 42show the component volumes
which contribute to system delay volume in the Agilent 1260 Infinity
Binary LC System .
Switching between configurations can be done in two ways:
•
manually, by disconnecting and reconnecting capillaries
•
automatically, using a 600 bar 2PS/6PT valve (optional)
Table 4
Delay volumes of the 1260 Infinity Binary LC modules
Components
Delay Volume (µL)
Binary Pump
1
1
in low delay volume configuration with bypassed damper and mixer
120
Binary Pump
2
2
in standard delay volume configuration
600 – 800
Low volume mixer
200
Mixer
400
Autosampler
270
Low dispersion heat exchanger
1.6
Built-in heat exchanger
3 and 6
Table 5
Delay volumes of 1260 Infinity Binary LC configurations
System Configuration
Delay Volume (µL)
Low delay volume configuration
Pump: 120
Autosampler:
Medium delay volume configuration
Pump: 320
Standard delay volume configuration
Pump: 600 – 800